276
Appendix
EVIS EXERA II VIDEO SYSTEM CENTER CV-180
Transportation, storage, and operation environment
Specifications
Transportation
and storage
environment
Ambient temperature 25 to +70C (13 to +158F)
Relative humidity 10 - 90%
Atmospheric pressure 700 - 1060 hPa
Operation
environment
Ambient temperature 10 - 40C (50 - 104F)
Relative humidity 30 - 85% (without condensation)
Atmospheric pressure 700 - 1060 hPa
(0.7 - 1.1 kgf/cm
2
)
(10.2 - 15.4 psia)
Item Specifications
Power supply Voltage 100 - 240 V AC
Voltage fluctuation Within 10%
Frequency 50/60 Hz
Frequency
fluctuation
Within 1Hz
Consumption
electric power
150 VA
Fuse rating 5 A, 250 V
Fuse size ø5 20 mm
Size Dimensions 382 (W) 91 (H) 490 (D) mm
(maximum)
Weight 10 kg
Classification
(medical electrical
equipment)
Type of protection
against electric
shock
Class I
Degree of
protection against
electric shock of
applied part
Depend on applied part
See also applied part (camera head
or videoscope).
Degree of
protection against
explosion
This instrument should be kept
away from flammable gases.