Functions
6.8 [2.09] Settings > Inputs/outputs
Operating with the Local User Interface
142 Operating Manual, 06/2017, A5E31930478-05
Setting the analog outputs
Procedure
Note
Component-specific setting
An analog output is available for each component. Repeat the procedure described below for
all components, i
f necessary.
1. Component > Main menu > "2. Settings" > "09. Inputs/outputs" > "2. Analog output"
The parameter display is opened.
2. Set the analog output range: "Analog output range" parameter box.
– Authorize access with Standard PIN, if necessary.
– Select one of the available list entries according to Table 6-15 Analog output range
and measuring range limit (Page 141).
3. Permit or suppress negative values: "Negative values" parameter box.
If negative measured values have a negative impact on the measured value processing,
select "Suppress".
Negative measured values are then set to 0 or 4 mA at the analog output while the
display still shows the negative measured value.
4. Set the behavior of the analog output with enabled function check: "At function check"
parameter box.
By setting this parameter you determine which value is output by the analog output on a
change to "Function check" operating state. The following values can be output as an
alternative:
Low limit of the analog output range
High limit of the analog output range
Last measured value before entering function check
5. Set the behavior of the analog output in case of a maintenance alarm or fault: "At
main.alarm/fault" parameter box:
Setting range and setting options correspond to step 4.
Overview (Page 141)
Setting the measured-value format (Page 98)