Functions
6.9 [2.10] Settings > Correction of cross-interferences
Operating with the Local User Interface
172 Operating Manual, 06/2017, A5E31930478-05
Assigning correction of cross-interference with a constant
Procedure with ULTRAMAT 7
1. Component > Main menu > "2. Settings" > "10. Correction of cross interference" > "2
Constant"
2. Enable/disable correction of cross-interference: "Correction of cross-interference"
parameter field.
Authorize access with Expert PIN, if necessary.
3. Enter constant already known: "Constant" parameter field. The measured value is
corrected by the amount entered.
4. If necessary, change assignment of process tags for correction of cross-interference
Note
Setting instruction
You can only change the assignment of process tags for correction of cross
-interference
if process tag switching is enabled. To do this, proceed as follows:
Check the displayed process tag assignment.
"Process tag" > "Change" parameter field
The displayed editing window includes all available process tags.
Activate the corresponding process tags that are to be integrated into the correction of
cross-interference. Deactivate process tags that are not used.
5. Enabling/disabling all corrections of cross-interferences (Page 179).
Assigning variable correction of cross-interference
Procedure with ULTRAMAT 7
1. Component > Main menu > "2. Settings" > "10. Correction of cross-interference" > "3 ... 6
..."
2. Enable/disable correction of cross-interference: "Correction of cross-interference"
parameter field. Authorize access with appropriate Expert PIN, if necessary.
The correction of cross-interference is activated/deactivated for a specific correction
value.