Description
3.6 Combined operation
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
45
Requirement for successful module retrofitting or successful module replacement
The firmware of the module to be replaced must be compatible with the firmware of the
processing module (PM).
Installation of a new module in an existing analyzer:
● You must update the firmware of the PM prior to installation of the module.
If an analyzer module is already present, you must update its firmware.
Option modules (Page 30)
Plug & measure permitted activities (Page 44)
Case differentiations (Page 46)
Damage caused by inadmissible procedure/faulty operations
Inadmissible procedure and/or faulty operations can result in damage to the device or its
modules/electrical equipment in connection with Plug & Measure. Therefore:
• Observe the specifications regarding installation/removal of the device hardware.
• Carefully read the following information on different Plug & Measure cases.
Damage to the analyzer module electronics due to faulty operation during commissioning
During commissioning after replacement of the AM electronics, a faulty operation can
prevent transfer of the parameters and render the AM electronics unusable. When you
commission the device once again after replacing the AM electronics, the following applies:
• First, confirm the message regarding change of the configuration
• Make sure to confirm the transfer of the parameters by pressing <ENTER>