Functions
6.1 [1] Quick Start
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
91
4. Enable/disable measuring range: "Use measuring range" parameter field.
Note
Using measuring ranges in other device functions/menus
Other device functions or menus can only access measuring ranges activated in this
menu. Inactive measuring ranges are not displayed for selection in oth
er device
functions/menus. This means they can no longer be used, for example, in autoranging or
in calibrations.
5. Set additional measuring ranges/parameters, e.g., → Setting Autoranging (Page 91).
Setting the setpoints (Page 121)
Displaying current messages (Page 203)
Setting Autoranging
Procedure
Note
Autoranging
Autoranging is not possible if the measuring ranges were set incorrectly. In this case
measuring range 1 is
selected as active measuring range.
Setting the measuring ranges (Page 89)
Note
Component-specific setup
Measuring ranges are comp
onent-specific. Repeat the procedure described below for all
components, if necessary.