Functions
6.12 [2.13] Setting > Process tag switchover
Operating with the Local User Interface
186 Operating Manual, 06/2017, A5E31930478-05
1. Main menu > "2. Settings" > "13. Process tag switchover" > "02. ..." to "13. ..."
2. Specify the sequence of the process tags within the automatic process tag switchover:
"Move position to" parameter field
– Authorize access with Standard PIN, if necessary.
– Enter the required position number (1 to 12).
3. Activate/deactivate process tags for automatic process tag switchover:
"Include process tag" parameter field (on/off).
4. Specify the name of the process tag: "Designation of process tag" parameter field.
You can use up to seven characters for the process tag designation.
5. Check link with digital outputs: "Process tag combined with" display field.
Change the assignment of digital outputs, if necessary: Setting digital outputs (Page 155).
6. Set the measuring time: "Measuring time" parameter field.
Set the time for which the process tag is active during the switchover cycle.
7. Read calculated remaining time: "Remaining time" display field
If the process tag is inactive, the remaining time is the same as the assigned measuring
time.
If the process tag is active, the remaining measuring time until the next switchover is
displayed.
8. Activate automatic process tag switchover: Enabling/disabling all process tags
(Page 186).
Enabling/disabling all process tags
Procedure
1. Main menu > "2. Settings" > "13. Process tag switchover" > "01. Total"
This function gives you the option to switch off the process tag switchover and to switch it
on again at one location.
2. Enable/disable all process tags for process tag switchover:
"Process tag switchover" parameter field (on/off)
– "Off": When you select this setting, the currently active process tag remains
permanently active.
– "On": The process tag switchover starts immediately with the process tag at position 1.
Only process tags already enabled under Assigning process tags (Page 185) are
considered during the switchover.