Functions
6.16 [3.02] Maintenance & Diagnostics > Messages to be acknowledged
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
207
6.16.2
Acknowledge message
Procedure
1. Main menu > Main menu > "3. Maintenance & Diagnostics" > "2. Messages to be
acknowledged" > "1. Messages to be acknowledged"
This menu includes all messages to be acknowledged that match your parameter
assignment under [2.14] Settings > Set message parameters (Page 187). The menu can
have hundreds of entries. The status bar also includes the current date and the current
time. All messages to be acknowledged are assigned instructions for remedial measures.
The menu entries have the following structure:
Structure of the navigation lines
1)
2)
Device
3)
0208
4)
Ext. maint. demand 1:
5)
Date
6)
Time
6)
Shortened message text....
â–º
1)
2)
Device
3)
0209
4)
Ext. maint. demand 2:
5)
Date
6)
Time
6)
Shortened message text....
â–º
1)
Symbol for identification of measuring type
2)
Acknowledgment status. Message not acknowledged:
3)
Message reference: message refers to the device, a module, or a component.
4)
Message number
5)
Message name
6)
Date / time: shows the "Time incoming".
2. The following information is shown when you open a message:
– Description of cause
– Appropriate remedial measures
– "Acknowledge" parameter field.
– Time incoming
3. Remedy the cause of the message:
The list entry is only deleted in the menu [2.14] Settings > Set message parameters
(Page 187) if you acknowledge the message before and after you remedy the cause.
4. Acknowledge message: "Acknowledge" parameter field.
– Main menu > "3. Maintenance & Diagnostics" > "2. Messages to be acknowledged" >
"1. Messages to be acknowledged" > " ... "
– This function is triggered immediately after the Standard PIN is entered.

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIPROCESS GA700 and is the answer not in the manual?

Siemens SIPROCESS GA700 Specifications

General IconGeneral
BrandSiemens
ModelSIPROCESS GA700
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals