Functions
6.15 [3.01] Maintenance & Diagnostics > Current messages
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
205
6.15.3
Reading current messages
Procedure
1. Main menu > "3. Maintenance & Diagnostics" > "01. Current messages" >
This menu includes all messages that match your parameter assignment under → [2.14]
Settings > Set message parameters (Page 187). The menu can have hundreds of entries.
The status bar also includes the current date and the current time. All current message
are assigned instructions for remedial measures.
The menu entries have the following structure:
Structure of the navigation lines
1)
2)
Device
3)
0208
4)
Ext. maint. demand 1:
5)
Date
6)
Time
6)
Shortened message text....
â–º
1)
2)
Device
3)
0209
4)
Ext. maint. demand 2: Date
6)
Time
6)
Shortened message text....
â–º
1)
Symbol for identification of measuring type
2)
Display of acknowledgment status. acknowledged, yes: / No:
An empty column indicates that the message does not require acknowledgment.
3)
Message reference: message refers to the device, a module, or a component.
4)
Message number
5)
Message name
6)
Date / time: The time information reflects the "Time incoming" of the message.
2. The following information is shown when you open a message:
– Description of cause
– Appropriate remedial measures
– Time incoming
3. Remedy the cause of the message:
– Messages that do not require acknowledgment:
When you implement the remedial measures, the message is no longer present in the
list of current messages.
– Messages that require acknowledgment:
The list entry is only deleted if you acknowledge the message before and after you
remedy the cause.
4. If necessary, change the filter settings: → Filtering current messages (Page 204).

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIPROCESS GA700 and is the answer not in the manual?

Siemens SIPROCESS GA700 Specifications

General IconGeneral
BrandSiemens
ModelSIPROCESS GA700
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals