Alarms and Messages
VF4 Infinity Gamma Series Page 5-9
Turning Alarm Recordings On/Off
The monitor prints and/or stores alarm recordings automatically,
if the Record and/or Store function is enabled on the Alarm Lim-
its table.
STEPS: Turning Alarm Recordings/Storage On or Off
1. Press the Alarm Limits fixed key.
2. Scroll to the Record column of the desired parameter and
click the knob.
3. Dial in the desired recording/storage setting and click the
knob.
See the Recordings chapter for a description of alarm recordings,
stored recordings, and the Event Recall screen.
External Alarm Lights
A set of alarm lights on top of the monitor blink red for life-
threatening alarms and yellow for serious alarms, if the external
alarm light function is enabled in the Monitor Setup menu
(default setting is ON). If more than one alarm occurs at the same
time, the lights blink for the alarm with the highest alarm grade. If
the user silences the alarm, the alarm lights remain lit without
flashing. If the user turns all alarms off, the alarm lights are
turned off as well.
For important information about external alarm lights, see the
chapter Monitor Setup.

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
BrandDräger
ModelInfinity Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals