End-Tidal CO2
VF4 Infinity Gamma Series Page 13-19
etCO2 Alarms
The monitor alarms for the following etCO
2
parameters:
! etCO
2
alarm (critical levels of CO
2
saturation in the exhaled
breath).
! iCO
2
alarm (critical levels of CO
2
saturation in the inhaled
breath, or rebreathing).
! RRc alarm (significant changes in the respiratory rate,
derived from the capnogram waveform).
! RRc Apnea alarm (apnea conditions based on the respiratory
rate derived from the capnogram waveform).
In addition, the monitor alarms for unplugged cables, weak sig-
nals, and in the absence of periodic measurement updates.
You can set etCO
2
upper and lower limits for all monitoring
modes within the following range:
! 5 to 95 mmHg for the etCO
2
alarm (0.7 to 12.7 kPa or %).
! 4 to 10 mmHg for the iCO
2
alarm (0.5 to 1.3 kPa or %).
! 5 to 145 breaths per minute for the RRc alarm (mainstream).
! 5 to 65 breaths per minute for the RRc alarm (sidestream).
Turn etCO
2
alarms on or off and set the alarm limits in the Alarm
Limits table (see the chapter Alarms and Messages).

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
BrandDräger
ModelInfinity Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals