EasyManuals Logo
Home>Dräger>Medical Equipment>Infinity Series

Dräger Infinity Series User Manual

Dräger Infinity Series
384 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #324 background imageLoading...
Page #324 background image
Page B-12 Infinity Gamma Series VF4
Cleaning/Replacing the Fan Filter
The fan filter should be cleaned once per month and replaced at
least once a year.
STEPS: Cleaning/Replacing the Fan Filter
1. Grasp the fan filter and remove it from its holding slots on the
back of the module.
2. Vacuum up any accumulation of dust at the fan port and
inside the filter (if reusing the filter).
3. Reinsert the cleaned fan filter or a new filter.
Fan Filter

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
BrandDräger
ModelInfinity Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals