Recordings
VF4 Infinity Gamma Series Page 7-11
Recording Trends
Trend printouts are long-term records of heart rate and other
patient vital signs. The printouts are snapshots of the tableor
graphs as displayed when you press the
Record fixed key on the
monitor.
Trend parameters are printed in the order selected by the user. In
addition, parameter trend data is printed only if valid trend data is
available for that parameter in the tabular trend columns or in the
graphical trend window.
You must view the graphical and tabular trends in order to record
them. An alarm recording request preempts a trend recording.
In monitors that display Neonatal OCRG, the recorder prints 144
seconds of patient OCRG data that occurred prior to pressing the
Record fixed key. The illustration below shows an OCRG
recording strip.