Cleaning, Disinfecting, Sterilizing
VF4 Infinity Gamma Series Page B-11
Scio Module and Accessories
WARNINGS:
! Occupational safety hazared: Used sampling tubing,
T-connectors, and water traps could be contaminated
and must be handled and disposed of with care.
Infection hazard may be present. Dispose of these
items in accordance with local regulations.
! Because of the danger of electric shock, never
remove the cover of any device while it is in opera-
tion or connected to a power outlet.
Emptying the Water Trap
The water trap should be emptied if the contents has reached the
‘full’ mark. To empty the water trap:
1. Disconnect the sampling tubing.
2. Remove the trap from its receptacle by
holding it firmly on the ridged surfaces and
pulling it out from the Scio module.
3. Connect an empty syringe (size > 20 ml
without a needle) to the port on the back of
the water trap.
4. Pull water trap contents into the syringe.
5. Remove syringe and discard.
NOTE: Do not clean the water trap. If the water trap seems dirty or
if it has been in use for 4 weeks, replace it.