Page 8-10 Infinity Gamma Series VF4
Cascade Display
The electrocardiogram displayed in the first channel is about four
seconds long. To display eight seconds of the same waveform,
select the cascade mode. Cascade mode continues the display of
the first waveform channel into the second channel.
NOTES:
! The Cascade is only available in the second waveform channel.
If etCO2, IBP2 and ST are enabled, the ECG cascade is not
available.
! The second channel of a cascaded waveform cannot be printed
on recordings.
STEPS: Selecting the Cascade Display
1. Click on the second waveform channel.
2. Click on
Waveform.
3. Select
Cascade and click the knob.
NOTE: You cannot adjust the display amplitude (Size) for the sec-
ond waveform channel when Cascade is selected; the amplitude is
the same as the one in the first channel.