Page 14-2 Infinity Gamma Series VF4
Overview
NOTES:
! Multigas monitoring is only available for the monitor
Gamma XL (not for the monitor Gamma). Multigas monitor-
ing is a locked option and restricted to those clinical sites that
have the corresponding gas monitoring capabilities and deliv-
ery systems in place.
! Multigas monitoring is only available in the Adult and Pediat-
ric monitoring modes.
! During Multigas monitoring, the monitor does not display the
ST parameter box and ST monitoring is disabled.
The Gamma XL supports multigas monitoring functions when
operating in the OR mode. The monitor receives multigas data
from a Dräger Scio multigas module and can display concentra-
tions of O
2
, CO
2
, N
2
O, and of the anesthetic agents halothane,
isoflurane, enflurane, sevoflurane, and desflurane.