Technical Data
VF4 Infinity Gamma Series Page D-9
Monitoring Accessories
etCO2 Pod
Size (H x W x D) Pod: approx. 140 x 140 x 51 mm (5.5 x 5.5 x 2 in)
Capnostat
TM
III Sensor (without cable):
33 x 42 x 22 mm (1.3 x 1.7 x 0.9 in.)
Weight Pod: 1.1 lb or .5 kg
Capnostat
TM
III Sensor: 18 g
Connections Sensor connector, female luer side-stream sam-
pling port, male luer sample exhaust port
Adult Airway Adapter
Dead Space
< 5 cc
Neonatal Airway Adapter
Dead Space
< 0.5 cc
Moisture Resistance Airway adapter can be immersed in water without
damage
Biotoxicity None per ASTM F720 (MTL #2009)
Power Source Powered directly from monitor
Protection Against
Electric Shock
Type CF (per IEC 601-1)
Mode of Operation Continuous
Defibrillation Protection per IEC 601-1 A2
Mechanical Shock 20 one-meter drops per IEC 68-2-32 (1990), EN
60068-2-32 (1993)
Environmental
Requirements
(Readings comply to BTPS = Body
temperature of 37°C, ambient baro-
metric pressure of 750 mmHg, and
relative humidity of 100%.)
Temperature:
(operating:) 10 to 40 °C (50 to 104 °F)
(storage:) -20 to 50 °C (-4 to 122 °F)
Relative Humidity:
(operating:) 20 to 90%, non-condensing
(storage:) 10 to 95% (with packaging)
Atmospheric Pressure:
(operating:) 525 to 795 mmHg (70 to 106 kPa)
(storage:) 375 to 795 mmHg (50 to 106 kPa)