EasyManuals Logo
Home>Dräger>Medical Equipment>Infinity Series

Dräger Infinity Series User Manual

Dräger Infinity Series
384 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #353 background imageLoading...
Page #353 background image
Technical Data
VF4 Infinity Gamma Series Page D-9
Monitoring Accessories
etCO2 Pod
Size (H x W x D) Pod: approx. 140 x 140 x 51 mm (5.5 x 5.5 x 2 in)
Capnostat
TM
III Sensor (without cable):
33 x 42 x 22 mm (1.3 x 1.7 x 0.9 in.)
Weight Pod: 1.1 lb or .5 kg
Capnostat
TM
III Sensor: 18 g
Connections Sensor connector, female luer side-stream sam-
pling port, male luer sample exhaust port
Adult Airway Adapter
Dead Space
< 5 cc
Neonatal Airway Adapter
Dead Space
< 0.5 cc
Moisture Resistance Airway adapter can be immersed in water without
damage
Biotoxicity None per ASTM F720 (MTL #2009)
Power Source Powered directly from monitor
Protection Against
Electric Shock
Type CF (per IEC 601-1)
Mode of Operation Continuous
Defibrillation Protection per IEC 601-1 A2
Mechanical Shock 20 one-meter drops per IEC 68-2-32 (1990), EN
60068-2-32 (1993)
Environmental
Requirements
(Readings comply to BTPS = Body
temperature of 37°C, ambient baro-
metric pressure of 750 mmHg, and
relative humidity of 100%.)
Temperature:
(operating:) 10 to 40 °C (50 to 104 °F)
(storage:) -20 to 50 °C (-4 to 122 °F)
Relative Humidity:
(operating:) 20 to 90%, non-condensing
(storage:) 10 to 95% (with packaging)
Atmospheric Pressure:
(operating:) 525 to 795 mmHg (70 to 106 kPa)
(storage:) 375 to 795 mmHg (50 to 106 kPa)

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
BrandDräger
ModelInfinity Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals