Page 5-12 Infinity Gamma Series VF4
If the monitor detects two or more new alarm conditions dur-
ing the silence period:
! The monitor delivers the audible signal for the newest alarm
with the highest priority.
! Both alarming parameter boxes blink.
! The message for both alarms is displayed alternately in the
message area of the screen.
All Alarms OFF Key
Press the All Alarms OFF fixed key to suspend all alarm indica-
tions for a 3 minutes. During an alarm suspension, the message
banner All Alarms OFF appears on top of the waveform area.
NOTE: When starting the monitor, coming out of Standby, or
answering “Yes” to the New Patient prompt, alarms are suspended
for 3 minutes or until you press the All Alarms OFF key to deacti-
vate the alarm suspension.
During the 3-minute alarm suspension:
! All currently annunciated alarm tones cease.
! All parameter boxes return to normal colors.
! All alarm messages are removed from the message area at the
bottom of the display.
! All new alarms are blocked.
! Currently activated alarms, including latched alarms, are
acknowledged.
! When on the network, the message banner All Alarms OFF
appears in the bedside’s remote display on the M
ULTIVIEW
W
ORKSTATION and in the ClusterView.
WARNIN G: During alarm suspensions, never leave a
patient unattended, and always re-enable the alarms as
soon as possible.