EasyManuals Logo

Fresenius Medical Care multiFiltrate User Manual

Fresenius Medical Care multiFiltrate
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #100 background imageLoading...
Page #100 background image
Chapter 4: Operation
4-58 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Paediatric CRRT treatments
4.7.6.5 Connecting the patient
Tip
If the patient is not yet available, then recirculation can be continued by
pressing the [START/RESET] key.
Note
In the case of CVVH, connect the substituate line in predilution,
if necessary.
If using an NaCl solution with two
connectors:
 Disconnect the arterial patient line from
the NaCl solution and connect it to the
arterial vascular access.
or
 Connect the arterial and venous patient
line to the vascular access.
If using an NaCl solution with one
connector:
 Connect the arterial and venous patient
line to the vascular access.
and
 Use the rotary selector to select Start
connection? [OK] to confirm! and press
[OK].
The blood pump will deliver at the
programmed rate.
Settable rate:
Paediatric 10 ml/min to 50 ml/min
(default 50 ml/min)