Chapter 4: Operation
4-40 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
CRRT treatments with citrate anticoagulation
4.6.6.3 Priming the tubing system
4.6.6.4 Rinsing the tubing system / entering treatment parameters
 Use the rotary selector to select Start
priming? [OK] to confirm! and press
[OK].
Heparin can be added to the NaCl solution
used for priming and rinsing via the Bolus
menu field.
 Open the infusion / extraction point to
prime the arterial bubble catcher.
If the amount of air in the arterial bubble
catcher has decreased to a level of approx.
1 cm underneath the lid, the infusion /
extraction point has to be closed.
Warning
Please ensure that the ratios of blood flow to dialysate flow / substituate
flow are properly set.
Note
Please set the treatment parameters (citrate dose, calcium dose, blood
flow, dialysate flow) as described in chapter 7.