Chapter 9: Installation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 9-1
9Installation
9.1 Connection requirements
9.1.1 Environment
The following considerations need to be taken into account for the
operating environment:
– No splash water area
– Ceilings, walls, floors: smooth, liquid-tight, scrub-resistant, suitable
for wet disinfection
– Ensure adequate load-carrying capacity of the floors
– Heat emission of each device (dialysis operating mode: approx.
550 W at 20 °C room temperature)
– Space requirements of each device approx. 1 m
2
– Emergency lighting (for at least 1 hour in case of power failure)
– Distances to areas such as MRI scanner rooms
9.1.2 Power supply network
Power supply network requirements:
– The requirements specified by IEC 60364-7-710 for Group 1 rooms
must be met.
– Power failures < 20 ms
– An earthing system must be installed as prescribed.
– A power socket with a protective earth conductor connection is
required.
– The line cross-section and the line lengths to the wall outlet must
ensure that the voltage tolerance and the function of the protective
devices is always guaranteed. Recommended line cross-section to
the power socket: at least 3 x 1.5 mm
2
copper core for 220 V–240 V
and at least 3 x 2.5 mm
2
copper core for voltages of less than 220 V.
– Each electric circuit is protected from damage through fault
conditions with an automatic, fast-acting circuit-breaker
(recommendation: 16 A for 220 V–240 V and 20 A for voltages of
less than 220 V).
– No more than 1 device per wall outlet and electric circuit.
– The use of multiway sockets or extension cables is prohibited.
– Residual-current devices (RCDs) which protect against dangerous
shock currents in the event of fault conditions. One residual-current
device (RCD less than 30 mA) for each device or electric circuit.
– Overvoltage / lightning protection in the main and emergency power
supply networks.
– A connecting bolt must be available for an additional equipotential
bonding conductor.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care multiFiltrate and is the answer not in the manual?

Fresenius Medical Care multiFiltrate Specifications

General IconGeneral
BrandFresenius Medical Care
ModelmultiFiltrate
CategoryMedical Equipment
LanguageEnglish

Related product manuals