Chapter 4: Operation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-29
CRRT treatments
4.5.9 Change of procedure back to CVVHDF
 Use the rotary selector to select
Substituate and set it to Off.
Warning
The storage life for opened solution bags indicated by the manufacturer
must be observed. If the storage life has been exceeded, the solutions
must not be used anymore.
Substituate or dialysate must remain deactivated or the end of the
treatment has to be initiated.
 Use the rotary selector to select
Substituate or Dialysate and set the
required flow.