Chapter 5: Alarm processing
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 5-11
5.9 Handling alarm limits in the event of an alarm
A Pressure alarm is displayed.
TMP, arterial, or venous
The actual value is outside the alarm limits.
Example: Arterial pressure.
The alarm limits are flashing.
The pressure display is red.
The blood pump is stopped.
 Use the rotary selector to select the green
question mark and press [OK].
The possible causes will be displayed.
 Correct the condition causing the alarm.
 Press the [START/RESET] key.
The blood pump will start at a slower
speed and gradually increase to the
prescribed rate. Then the required
balancing pumps will start (depending
on the treatment mode).