Chapter 4: Operation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-35
CRRT treatments with citrate anticoagulation
4.6.3 Inserting further tubing systems
Depending on the selected treatment mode, the tubing system for
dialysate and / or substituate has to be inserted.
4.6.3.1 Inserting the dialysate tubing system
 Insert the citrate and calcium tubing
system according to the instructions.
 Turn the rotary selector clockwise until the
next screen displays.
Warning
Please ensure that the dialysate / substitution fluid used is suitable for
the selected Ci-Ca treatment.
Note
Depending on the settings, the dialysate and substituate tubing
arrangement may differ from the illustration on the screen. The device
has to be set up in accordance with the instructions on the screen.
The settings for the tubing system may only be changed by technical
service support.
Warning
Ci-Ca CVVHD and Ci-Ca postCVVHDF treatments may only be
performed if calcium-free dialysate is used.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care multiFiltrate and is the answer not in the manual?

Fresenius Medical Care multiFiltrate Specifications

General IconGeneral
BrandFresenius Medical Care
ModelmultiFiltrate
CategoryMedical Equipment
LanguageEnglish

Related product manuals