Chapter 12: Specifications
12-12 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Rinse / recirculation Rinse volume: 0 ml to 5000 ml, can be set in the System parameters
menu
UF rinse: 300 ml to 2000 ml, can be set in the System parameters menu
Treatment SCUF, CVVH, Pre-Post CVVH, CVVHDF, CVVHD, MPS, and HP
Bag change Minimum plasma volume 50 ml
End of treatment /
reinfusion
Reinfusion volume: 0 ml to 2000 ml, can be set in the System
parameters menu
System parameters After the functional test and selection of the therapy, the audible alarm
volume, the display brightness, the key sounds, and the default values
for the treatment can be defined in the Setup menu at any time.
Selection of the treatment modes is inhibited whilst treatment is in
progress. It is only possible to select the blood line system and to set
date and time before priming / rinsing starts.
Audible alarm Adjustment range of the sound pressure level in the System parameters
menu is 65 dBA (20%) to 84 dBA (100%).
12.10 Balancing / dialysate circuit and safety systems
Blood leak detector Response threshold smaller than or equal to 0.5 ml blood loss per
minute at a haematocrit of 32% at maximum filtrate flow.
Optical adsorption system (red / green ratio).
The response threshold is related to the maximum filtrate flow. The
initiation of a blood leak alarm also depends on the ultrafiltration rate
and the size of the membrane rupture in the filter.
Flow rates Depending on the treatment mode:
The delivery accuracy of the replaceable pumps is ± 10%, unless
regulated by the scales. If regulated (for treatment procedures with
scale balancing), the individual delivery accuracy of each pump
depends on the accuracy of the associated scale. In this case, the total
delivery accuracy corresponds to the specified balancing accuracy.
Balancing / dialysate
circuit
< 1% related to the total volume delivered during 24 hours of treatment,
in horizontal position and with a maximum inclination of 5°. A deviation
of ±100 ml is possible if the total volume delivered is low (typically up to
5000 ml).
Blood flow* 10 ml/min to 500 ml/min ±10%
Substituate flow* 10 ml/min to 80 ml/min, controlled
Dialysate flow* 10 ml/min to 80 ml/min, controlled
Plasma exchange rate* 10 ml/min to 50 ml/min, controlled
Ultrafiltration rate* 0 ml/min to 30 ml/min, controlled
Filtrate rate 10 ml/min to 180 ml/min, controlled

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care multiFiltrate and is the answer not in the manual?

Fresenius Medical Care multiFiltrate Specifications

General IconGeneral
BrandFresenius Medical Care
ModelmultiFiltrate
CategoryMedical Equipment
LanguageEnglish

Related product manuals