Chapter 4: Operation
4-48 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
CRRT treatments with citrate anticoagulation
4.6.9 Disconnecting the patient and removing the tubing system
Indication of the decreasing reinfusion
volume.
Selecting View treatment history? [OK] to
confirm! will display the treatment
parameters.
The optical detector senses non-opaque
fluid.
 When selecting Continue reinfusion, the
remaining reinfusion volume can be
returned to the patient.
or
 Use the rotary selector to select
Terminate reinfusion and press [OK].
The citrate pump stops.
but
The treatment will be terminated
automatically if the reinfusion volume is 0 ml
and the optical detector senses non-opaque
fluid.
The citrate pump stops.
Warning
It is forbidden to remove the Ci-Ca lines manually before the patient is
disconnected.
Warning
Consumables must be discarded after the treatment in compliance with
the regulations for the disposal of potentially contaminated materials.