Chapter 4: Operation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-39
CRRT treatments with citrate anticoagulation
4.6.6.2 Priming the Ci-Ca lines
The Ci-Ca pumps fill the Ci-Ca lines with fluid
for a defined time period.
In that time period, entries cannot be
made on the multiFiltrate. Citrate
anticoagulation cannot be deselected
before the Ci-Ca lines are filled
completely.
The filling level of the Ci-Ca lines can be
readjusted.
 Use the rotary selector to select "Refill
citrate" or "Refill calcium" and press
[OK].
The particular pump will run for one
revolution. This procedure can be
repeated as often as required.
 Once the Ci-Ca lines are primed while
being free from air, use the rotary selector
to select Ci-Ca lines primed and free
from air? [OK] to confirm! and press
[OK].