Chapter 4: Operation
4-28 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
CRRT treatments
4.5.8 Change of procedure (except for CVVHD and CVVH)
4.5.8.1 Change of procedure to CVVH
4.5.8.2 Change of procedure to CVVHD
Note
It is possible to stop a flow during the treatment in the CVVHDF and
Pre-Post CVVH treatment modes. Only a CVVH or CVVHD will then be
performed in the CVVHDF, for example. This change of procedure can
be undone.
The changed treatment mode is shown in the status bar. The letter
which is greyed out indicates a change of procedure.
 Use the rotary selector to select Dialysate
and set it to Off.
Note
"Sub. bolus 100 ml" is not possible in CVVHD.