Chapter 2: Important information
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 2-15
The alarm limits of the patient leakage current for type CF applied parts
are as follows:
– 10 µA AC / DC (normal case, error-free)
– 50 µA AC / DC (the first time an error occurs)
Please address any queries to the local service support organisation.
2.12.5 Warnings relating to consumables and accessories
Warning
Consumables must be discarded after the treatment in compliance with
the regulations for the disposal of potentially contaminated materials.