Chapter 4: Operation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-45
CRRT treatments with citrate anticoagulation
4.6.7.2 Treatment main screen
5 minutes after the treatment has been
started, a note is displayed, accompanied by
a short audible signal. This is the prompt to
check the post-filter calcium concentration
now.
The main screen is displayed throughout the
entire treatment.
Depending on the treatment mode, the menu
field shows:
The current pressures (arterial, venous,
TMP)
The current flow rates (dialysate, substituate
in postdilution, blood flow, ultrafiltration)
Citrate / blood dose and citrate flow
Calcium / filtrate dose and calcium flow
Heparin
The status bar shows:
The treatment mode
The progression of the treatment time
The balance
Continuous anticoagulation on / off