Chapter 12: Specifications
12-4 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
12.4 Electrical supply
Line voltage 100 V AC, 50–60 Hz, 6.3 A
110 V AC, 50–60 Hz, 6.0 A
115 V AC, 50–60 Hz, 5.5 A
120 V AC, 50–60 Hz, 5.4 A
127 V AC, 50–60 Hz, 5.2 A
220 V AC, 50–60 Hz, 3.3 A
230 V AC, 50–60 Hz, 3.2 A
240 V AC, 50–60 Hz, 2.7 A
(Always go by the line voltage, frequency, and operating current
information specified on the identification label attached to the device
itself.)
Power supply (internal) +24 V / 18 V, battery-backed, derived from this voltage:
+5 V ±0.2 V
+12 V ±0.3 V
-12 V ±0.3 V
+24 V +1 V / –2 V
Power switch Main power switch, all-pole, simultaneous disconnection
Battery Lead-acid battery (maintenance-free), 18 V / ≥3Ah
Operating time approx. 15 minutes
Blood circuit only, without balancing, without heater