Chapter 12: Specifications
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 12-3
12.2.3 Identification label of the Ci-Ca module
The identification label shown is only an example. Always go by the
information shown on the identification label affixed to the Ci-Ca module
itself.
Legend
1 Manufacturer and year of manufacture
2 Serial number
3 Connected load
12.3 Electrical safety
Classification according to EN 60601-1, IEC 60601-1
Type of protection against
electric shock
Protection class I with internal power source
Applied part Depending on the treatment mode, the applied part comprises the
extracorporeal blood circuit, the dialysate, substituate, and plasma
circuits, and all components with a permanent, conductive connection
to these circuits.
Degree of protection
against electric shock
220 V / 230 V / 240 V
60 Hz, type BF applies
100 V / 110 V / 115 V / 120 V / 127 V
60 Hz, type CF applies
100 V / 110 V / 115 V / 120 V / 127 V / 220 V / 230 V / 240 V
50 Hz, type CF applies
Defibrillator-proof applied
part
A defibrillator-proof applied part prescribes the exclusive application of
disposables as approved by the manufacturer.
Samples of approved tubing systems are listed in chapter 8
(see chapter 8.1.1 on page 8-2).
Degree of protection
against ingress of liquids
Drip-proof, IP 21
Leakage currents According to EN 60601-1
1
3
2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care multiFiltrate and is the answer not in the manual?

Fresenius Medical Care multiFiltrate Specifications

General IconGeneral
BrandFresenius Medical Care
ModelmultiFiltrate
CategoryMedical Equipment
LanguageEnglish

Related product manuals