Chapter 4: Operation
4-70 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Membrane plasma separation (MPS)
4.8.6.3 UF rinse
4.8.6.4 Recirculation / waiting for patient
Note
When using NaCl solutions with only one connector, make sure there is
enough NaCl solution.
After the rinse is completed and the
pre-defined volume used, the blood pump will
stop.
An audible signal will be given.
If using an NaCl solution with two
connectors:
 Disconnect the venous patient line from
the empty bag and connect it to the NaCl
solution.
If using an NaCl solution with one
connector:
Do not change the existing connection.
 Use the rotary selector to select Start UF
rinse? [OK] to confirm! and press [OK].
Note
The filter life, which is shown in the status bar, is automatically added to
the service life of the tubing systems.
When the UF volume reaches 0 ml, the
extracorporeal blood circuit is in recirculation.
If using an NaCl solution with two
connectors:
Do not change the existing connection.
If using an NaCl solution with one
connector:
 Stop recirculation by pressing the [STOP]
key.
Press the [STOP] key for approx.
3 seconds.
 Connect the venous patient line to the
arterial patient line using the recirculator.
 Continue recirculation by pressing the
[START/RESET] key.