Chapter 4: Operation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-105
Treatment menu
4.11.2 Selecting Ci-Ca anticoagulation
 Use the rotary selector to select Yes from
the menu field and press [OK].
 Use the rotary selector to select All
conditions fulfilled? [OK] to confirm!
and press [OK].
After deselecting citrate anticoagulation, the
multiFiltrate immediately switches to the
menu for dialysate bag change.
Warning
Ci-Ca CVVHD and Ci-Ca postCVVHDF treatments may only be
performed if calcium-free dialysate is used.
Warning
When starting the treatment, define the post-filter calcium value. If the
ionised calcium has not decreased at this point, it is absolutely
necessary that the tubing system and the solutions used are checked.