Chapter 4: Operation
4-34 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
CRRT treatments with citrate anticoagulation
4.6.2 Inserting the cassette system
4.6.2.1 Inserting the citrate tube and the calcium tube
 Insert the cassette system according to
the instructions.
 Turn the rotary selector clockwise until the
next screen displays.
Warning
When inserting the citrate and calcium tubes, ensure that the tube
segments and reservoirs for the solutions concerned are properly
locked and assigned.
Warning
The citrate and calcium solutions must be at room temperature before
they can be used.
Note
The following requirements must be fulfilled before inserting the Ci-Ca
pump segments to be able to perform automatic priming:
– The clamps (white and green) are moved to the connectors.
– The clamps (white and green) are closed.