Chapter 4: Operation
4-30 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
CRRT treatments
4.5.10 End of treatment
4.5.10.1 Terminating the treatment
4.5.10.2 Starting reinfusion
 Press the [ESC] key.
 Use the rotary selector to select End of
treatment from the menu bar and press
[OK] to confirm.
The blood pump is running.
 Stop the treatment by pressing the
[STOP] key.
Press the [STOP] key for approx.
3 seconds.
The venous clamp closes.
The blood pump is stopped.
or
 Use the [ESC] key to select a different
menu from the menu bar.
 Continue the treatment by pressing the
[START/RESET] key.
or
 Disconnect the arterial patient line from
the patient and connect it to the NaCl
solution.
 Use the rotary selector to select Start
disconnection? [OK] to confirm! and
press [OK].
When selecting Disconnect without blood
return, the program will go directly to
"Disconnecting the patient" without reinfusion
after a safety prompt (see chapter 4.5.11 on
page 4-32).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care multiFiltrate and is the answer not in the manual?

Fresenius Medical Care multiFiltrate Specifications

General IconGeneral
BrandFresenius Medical Care
ModelmultiFiltrate
CategoryMedical Equipment
LanguageEnglish

Related product manuals