Chapter 4: Operation
4-76 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Membrane plasma separation (MPS)
4.8.8 End of treatment
4.8.8.1 Infusing remaining plasma
The plasma is returned via the Change bag
menu.
 Confirm the message by pressing the
[START/RESET] key.
 Use the rotary selector to select Plasma
return and press [OK].
 Use the rotary selector to select Start
plasma return? [OK] to confirm! and
press [OK].
The residual plasma of approx. 300 ml will be
administered until the Balancing message is
displayed.