Chapter 5: Alarm processing
5-18 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
5.14 Power failure (mains failure)
A Power failure is displayed.
Balance is off.
The blood pump is running.
Citrate pump continues running for up to
10 minutes (then alarm due to
switched-off balance).
The calcium pump is stopped.
An audible signal will be given.
When power is restored, the system will start
automatically.
End of emergency operation is displayed.
Balance is off.
The blood pump is stopped.
The extracorporeal blood volume must be
returned using a hand crank
(see chapter 5.15 on page 5-19).