Chapter 4: Operation
4-104 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Treatment menu
4.11 Treatment menu
4.11.1 Deselecting Ci-Ca anticoagulation
 Press the [ESC] key.
 Use the rotary selector to select
Treatment from the menu bar and press
[OK].
The menu fields shown depend on the
selected treatment mode.
 Use the rotary selector to select the
required parameter field from the menu
field and press [OK].
After a timeout, the display will automatically
return to the treatment main screen.
Warning
After deselecting citrate anticoagulation, the operator has to ensure an
alternative anticoagulation procedure.
Warning
After deselecting citrate anticoagulation, the CVVHD treatment may
only be continued / performed if a calcium-containing HF solution /
dialysate is used.
Warning
After deselecting citrate anticoagulation, the Ci-Ca lines must not be
removed from the pumps before the patient has been completely
disconnected.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care multiFiltrate and is the answer not in the manual?

Fresenius Medical Care multiFiltrate Specifications

General IconGeneral
BrandFresenius Medical Care
ModelmultiFiltrate
CategoryMedical Equipment
LanguageEnglish

Related product manuals