Chapter 10: Transport / storage
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 10-3
After relocating Establish the voltage supply and confirm the alarm by pressing the
[START/RESET] key.
Briefly perform the following checks:
– Check the connectors.
– Ensure that the filtrate bag hangs freely and does not touch any
other objects.
– Visually check the tubing systems (cassette) and the solution bags
for damage, leakage, and proper fit.
– Pay particular attention to "Application principles" chapter
(see chapter 4.1 on page 4-1).
10.2 Transport
In general, the device has to be transported without any lines inserted
and any load on the scales.
The device must not be carried. Use a lift, ramp, or similar to overcome
level differences.
If the device needs to be transported to a location that is not within the
immediate vicinity of its current location, then the relocation goes
beyond the scope of the previous section. In this case, the full initial
start-up procedure must be performed again at the destination.
When transporting the haemodialysis system in a vehicle, always
protect it with the appropriate packing materials and place it either
vertically or horizontally.
10.3 Storage
The device must be stored upright in a well-ventilated room with low
variations in temperature.
Maintenance of the built-in
battery
On receipt of the device, charge the battery as follows:
– Use the power cable to connect the device to the power supply.
– Actuate the power switch to turn the device on.
– Leave the device on for 10 hours.
If the device is not used, repeat this procedure every six months.
Note
To ensure that the internal battery is always charged and ready for use,
the device has to be supplied with line voltage and the power switch has
to be turned on.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care multiFiltrate and is the answer not in the manual?

Fresenius Medical Care multiFiltrate Specifications

General IconGeneral
BrandFresenius Medical Care
ModelmultiFiltrate
CategoryMedical Equipment
LanguageEnglish

Related product manuals