Chapter 4: Operation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-19
CRRT treatments
4.5.2.4 Inserting the filtrate line system
4.5.3 Inserting further tubing systems
Depending on the selected treatment mode, the tubing system for
dialysate, substituate, or sub predilution and sub postdilution has to be
inserted.
 Insert the filtrate line system according to
the instructions.
 Turn the rotary selector clockwise until the
next screen displays.
Note
Depending on the settings, the tubing arrangement for dialysate,
substituate, or sub predilution and sub postdilution may differ from the
illustration on the screen. The device has to be set up in accordance
with the instructions on the screen.
The settings for the tubing system may only be changed by technical
service support.
Tip
With the CVVHD and CVVH (not Pre-Post) treatment modes, both
scales can be loaded with solutions if 2 x HF female adapters are used
on 4 x HF male adapters.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care multiFiltrate and is the answer not in the manual?

Fresenius Medical Care multiFiltrate Specifications

General IconGeneral
BrandFresenius Medical Care
ModelmultiFiltrate
CategoryMedical Equipment
LanguageEnglish

Related product manuals