EasyManuals Logo

Fresenius Medical Care multiFiltrate User Manual

Fresenius Medical Care multiFiltrate
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #250 background imageLoading...
Page #250 background image
Chapter 12: Specifications
12-16 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Calcium pump* Pump type: roller pump
Pumping accuracy: pressure-dependent, ±10% (above typical pressure
range)
Delivery rate: 0; 1 ml/h to 100 ml/h (calcium / filtrate ratio: calcium
concentration per litre of filtrate: 0 mmol/l to 3 mmol/l)
While the Ci-Ca tube segments are threaded in / removed and the
Ci-Ca tubes are primed, the Ci-Ca pumps are running with higher
delivery rates.
(* = Important performance characteristics according to IEC 60601-1)
12.12 Materials used
Plastics CR neoprene
NBR (Buna-N)
PA 6.6 (polyamide)
PC Gfn (Makrolon)
PE (Hostalen)
POM (Delrin)
PS (polystyrene)
PVC tube, transparent
Silicone tube, transparent, reinforced
E-HGW 2372.1 (epoxy resin laminated plastic)
Cast resin WEVO PU 127F/30
PUR hard foam (BÜFADUR 1011/1151 with flame retardant)
Metals Al 99.5
AlCuMgPb
AlMg3
AlMgSi 0.5
AlMgSi 1
CuBe
SF-Cu hard, oxygen-free
CuZn (Ms 63)
St 37-2, 1.0035 galvanised
Steel, 1.4301
Steel, 1.4305
Spring steel, 1.4310
Electrical equipment Motors:
Copper, cast steel, cast zinc
Stepper motors with plastic gear
Plug connectors:
Copper, tin, plated gold, glass-fiber-reinforced thermoplastic
Transformers:
Copper, polyester / polyurethane, potting compound (PU 151/20),
UP resin, iron core

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care multiFiltrate and is the answer not in the manual?

Fresenius Medical Care multiFiltrate Specifications

General IconGeneral
BrandFresenius Medical Care
ModelmultiFiltrate
CategoryMedical Equipment
LanguageEnglish

Related product manuals