Chapter 4: Operation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-115
Treatment menu
4.11.8.2 Changing the venous width of the alarm limits window
4.11.8.3 Exiting the Alarm limits menu
Warning
Venous alarm limit
The lower venous alarm limit must be set as close as possible to the
actual venous pressure value. Changing the alarm limit window position
or width can adversely affect or even cancel out the efficiency of the
safety system in identifying extraneous blood loss.
 Use the rotary selector to select Venous
width and press [OK].
 Use the rotary selector to enter the
required window width and press [OK].
 Use the rotary selector to select Exit
Alarm limits menu? [OK] to confirm!
and press [OK].