Chapter 4: Operation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-51
Paediatric CRRT treatments
4.7 Paediatric CRRT treatments
General description of the paediatric CVVH and CVVHD treatments
with information on the differences between the individual treatments.
Make the device ready for operation (see chapter 4.4 on page 4-12).
4.7.1 Starting conditions
4.7.2 Inserting the AV set
4.7.2.1 Inserting the arterial blood line system
Note
The multiFiltrate paed CRRT / SCUF set tubing system has to be used
for paediatric CVVH and CVVHD treatment modes.
The status bar shows "Paed. CVVHD" or "Paed. CVVH" as the
treatment mode.
Note
Each treatment mode has its individual starting conditions.
The operator must ensure that the starting conditions are met before
using the rotary selector to select Conditions fulfilled and pressing
[OK].
 Insert the arterial blood line system
according to the instructions.
 Turn the rotary selector clockwise until the
next screen displays.