Chapter 4: Operation
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015 4-71
Membrane plasma separation (MPS)
4.8.6.5 Filling plasma
Tip
When using Fresenius fresh frozen plasma bags (FFP), for example,
the plasma pole (part no. M28 004 1) can be used if required.
 Stop recirculation by pressing the [STOP]
key.
Press the [STOP] key for approx.
3 seconds.
It is imperative to enter the net volume of the
replacement solution.
 Use the rotary selector to select Start
filling plasma? [OK] to confirm! and
press [OK].
The blood pump will deliver at the
programmed rate.
Settable rate:
10 ml/min to 100 ml/min (default
100 ml/min)
Time-controlled filling of plasma is in
progress.