CRRT treatments
4.5.6 Preparation
4.5.6.1 Priming the tubing system
This screen will be displayed immediately if
the setting in the System parameters menu
was set to Complete tubing arrangement.
 Confirm the completely inserted tubing
system using the rotary selector to select
Set up? [OK] to confirm! and press
[OK].
Note
The patient must not be connected during preparation. If the optical
detector senses opaque fluid during the preparation phase, the system
asks whether a patient has been connected.
 Use the rotary selector to select Start
priming? [OK] to confirm! and press
[OK].
Heparin can be added to the NaCl solution
used for priming and rinsing via the Bolus
menu field.