Chapter 4: Operation
4-122 Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Interrupting the treatment
4.13.1 Disconnecting the patient / interrupting the treatment
Note
– Make sure an NaCl solution with a volume of 1 l and two connectors
is available.
– The treatment should not be interrupted for more than 4 hours.
– Always observe aseptic procedures when connecting and
disconnecting the patient and perform the process with due care to
avoid any loss of blood.
 Set the blood flow to 100 ml/min and turn
the heparin pump off.
 Stop the treatment by pressing the
[STOP] key.
Press the [STOP] key for approx.
3 seconds.
 Disconnect the arterial patient line from
the patient and connect it to the NaCl
solution.
 Press the [START/RESET] key.
When the optical detector senses
non-opaque fluid, disconnect the venous
patient line from the patient and connect it to
the NaCl solution.
 Press the [START/RESET] key.
Balancing stops automatically.
Ci-Ca pumps stop automatically.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care multiFiltrate and is the answer not in the manual?

Fresenius Medical Care multiFiltrate Specifications

General IconGeneral
BrandFresenius Medical Care
ModelmultiFiltrate
CategoryMedical Equipment
LanguageEnglish

Related product manuals