Page #119 background imageLoading...
Page #119 background image
GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 3 - Setup Instructions 3-51
3-7-7-2 UP-D25MD - Printer Settings
To adjust the printer settings manually perform the following procedure:
1.) Call up the ‘Printer Preferences’; operation see: Section 3-7-7 "Adjustment of Printer Settings" .
2.) Select the
PAPER page and select:
- Paper Size: UPC-...L (large) / UPC-...S (small)
- Orientation: Landscape (recommended when using large paper size)
- High Speed (check mark on)
NOTE: Settings for paper size must match with the used paper(large/small) and also the right color ink
cartfridge has to be used. Otherwise you will get an error message at printing.
3.) Select the Graphics page. From the “Color Adjust” pull-down menu selet:
4.) Color Balance. Cyan=0, Magenta=0, Yellow=0
5.) Lightness. Sharpness = 7, Dark = 0, Gamma = -6, Gamma Curve = Curve 1
NOTICE
The appropriate printer settings can be automatically loaded by means of a “Registry file”. Therefore:
Download the SonyUp-D25MD.reg from GE folders
http://libraries.ge.com/foldersIndex.do?entity_id=15869258101&sid=101&SF=1
and extract files to CD or USB disk.
Connect the USB “standard GE” service dongle and enter the “Standard GE
Revolving” password.
Open the Windows Start menu by pressing CTRL +ESC on the keyboard, click on
“WinExplorer” and navigate to F:(where F stnads for CD/DVD Drive).
Double click the registry file.
Click YES to cinfirm.
Figure 3-51 Paper/Graphics page

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P6 BT18 and is the answer not in the manual?

GE Voluson P6 BT18 Specifications

General IconGeneral
Display15-inch LCD
Transducer Ports2
3D/4D ImagingYes
Hard Drive500 GB
Power Supply100-240 VAC, 50/60 Hz
Operating SystemWindows-based
Power Requirements100-240 VAC, 50/60 Hz
Storage Capacity500 GB
ApplicationObstetrics and Gynecology
Imaging ModesColor Doppler
Transducer Frequency Range2-9 MHz (depending on the transducer)
Probe TypesConvex, Linear, Endocavity, Volume
PortsUSB, Ethernet