GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-13
8-2-4 Software Update Package - Download/Installation Procedure
8-2-4-1 Overview
The system software installation procedure starts with saving and recording the settings present on the
system (silent "Full Backup" and "Rollback"). Then the new software is written to the hard disk.
Application Settings are automatically updated, to match with new Software version.
Existing User Programs, 3D/4D Programs and Auto Text remain unaffected! Afterwards the new
software is configured such that it is integrated again in its environment.
8-2-4-2 Update Procedure
1.) Remove any CD/DVD from the DVD drive.
2.) Perform Preparations as described in Section 8-2-1-1
3.) If not already done, disconnect all external USB devices (except DVD drive).
4.) Press the ON/OFF button on the control console.
5.) In the displayed screen click Download.
6.) Download takes some time; please wait. (Download maybe paused by means of the Pause button.)
This icon (at the right bottom of the screen) indicates that a new Software
Update Package is available for download and installation.
Figure 8-13 Download SW update package