GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 3 - Setup Instructions 3-79
3-12-2-10 WLAN Diagnostic
Running diagnostics is good if you think the adapterr might be bad. Diagnostics also tell you if the
connection is behaving properly. Sometimes the system connects by accident to the DVR or another
internal device which uses the TCP/IP protocol. The diagnostics would showw if there are errors when
performing the full two way communication. For example, if the IP address starts with 197 or 169 (
loopback addresses), then something is wrong.
1.) Press the “Utilities” key on the control console.
2.) On the right side of the screen select Connectivity and then click the Device Setup tab.
3.) Click the WLAN Configuration button; see Section Figure 3-64 on page 3-71
4.) The Wireless Network Configuration tool with available Wireless Networks appear.
5.) Select the Diagnostics tab and then click Run Diagnostics.
Figure 3-70 Run Diagnostics

Table of Contents