GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 3 - Setup Instructions 3-5
Section 3-3
Receiving and Unpacking the Equipment
The Voluson™ P8/Voluson™P6 ultrasound system and accessories are shipped from the factory in a
single durable shipping carton which is mounted on a raised wooden platform base.
* Weight is approximate and will vary depending upon the supplied peripherals
Before unpacking the unit:
Inspect the carton package for visible damage.
Inspect the Shock for evidence of accidental shock during transit, see: Figure 3-3 below).
CAUTION
!! CAUTION:
Please read this section carefully before unpacking the Voluson™ P8/Voluson™P6 and its
(optional) peripherals.
CAUTION
!! CAUTION:
Transport only with forklift or stacker truck.
During transport pay attention to the point of gravity (“tilt and drop” indicator)!
Have two people available to unpack the Voluson™ P8/Voluson™P6.
Attempts to move the unit considerable distances (or on an incline) by one person alone, could
result in personal injury, and/or damage to the system.
Table 3-4 Shipping Carton - Dimensions and Weight
Description Height Width Depth
Weight*
Voluson™ P8/Voluson™P6
incl. accessories
1490 mm 870 mm 650 mm
100kg (including
the system)
Figure 3-3 Shock Indicator
NOTICE
!! NOTICE:
The device must only be transported in the original packaging!
Each shipping crate is sealed. It is recommended to keep and store the shipping crate and all other
packing materials (including the support foams, anti-static plastic cover, etc.), in case the unit has to
be moved to a different location. Unpack the devices such a way that packaging can be reused.
For warranty purposes, storage of the above is required for one year from date of purchase..
NOTICE
!! NOTICE:
The device must only be transported in the original packaging!
If the shipping crate is damaged, please inform the Voluson™ P8/Voluson™P6 sales representative
immediately.
Indicator turns Red,
If the system got shock

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P6 BT18 and is the answer not in the manual?

GE Voluson P6 BT18 Specifications

General IconGeneral
Display15-inch LCD
Transducer Ports2
3D/4D ImagingYes
Hard Drive500 GB
Power Supply100-240 VAC, 50/60 Hz
Operating SystemWindows-based
Power Requirements100-240 VAC, 50/60 Hz
Storage Capacity500 GB
ApplicationObstetrics and Gynecology
Imaging ModesColor Doppler
Transducer Frequency Range2-9 MHz (depending on the transducer)
Probe TypesConvex, Linear, Endocavity, Volume
PortsUSB, Ethernet